Operating principle
In order to compare the partial pressures, the oxygen probe must be supplied with a gas of a known content of oxygen (ambient air : % 02 = 20.9 %). This is what we call reference air. Moreover, some of the Econox oxygen probes are equipped with a thermocouple allowing measurement of the actual treatment temperature.
Thus, the user who knows the CO content either by analysis or by theoretical calculation and who takes into account the values measured by the Econox oxygen probes, is able to determine the carbon potential (and the oxygen concentration) in the heat treatment atmosphere.
The measuring element of an oxygen probe has a limited lifetime. Its surface layer changes depending on the impurities which are found in the heat treatment atmosphere. Therefore the measuring element should be replaced when its impedance exceeds the tolerances admitted.
Cleaning air is used on our CarboProbe ZI pro and ZS Pro to allow cleaning of the zirconia measuring element with a flow of fresh air. Cleaning air is not available for the CarboProbe CP, DS, LT and HT.